Calibration Menu

To enter the

CALIBRATION MENU, press and hold the

and

buttons for 3 seconds.

Parameter

Set-Up Information

Press

To skip to next menu item

Press

To change sensor input sensitivity

This allows the calibration engineer to change the sensitivity range of the IHH200 when
connecting to sensors with a sensitivity of greater than 5mV/V. The IHH200 is factory set

for 5mV/V. To ensure the unit is set to 5mV/V press

To select 50mV/V you need to power down the unit and access the internal circuit board.
Move link LK1 and place it onto JP1. Power on the IHH200 and return to this point of the
calibration menu. You will notice that the menu parameter has changed to

SEnS 50.0

;

press

to change the sensitivity to 50mV/V and move on to the next parameter.

Press

to skip to next menu item

Press

to the set the display resolution

This parameter sets the decimal point position for the display and the resolution, i.e. a
value of

000.005

would display the reading to 3 decimal places and the readings will

change in steps of 0.005.

The decimal point position is moved one place to the right each time you press the

and

together.

Any value can be entered for the resolution, using the

and

arrows to select a

digit and the

and

arrows to increment or decrement the digits. Press

to

accept the value and move onto the next parameter.

To save the settings and move on to the next parameter press

THIS MENU IS DISABLED WHEN TEDS IS ENABLED

Press

to skip to next menu item.

Press

to the enter the calibration routine

If you have chosen to enter the calibration routine you will be prompted whether you want

to select

LiVE

. If you do not press

, press

. You will then be prompted to select

either of the other calibration methods, which in order, are

tAbLE

and

CAL

VAL.

To select

any of the calibration methods press

. Otherwise press

For more detailed calibration information, please refer to the calibration section of the
manual.