FUTEK IHH200 Digital Display User Manual

Page 21

Advertising
background image

TEDS limitations / specifications

ERROR 1

Must be DS2431 or DS2433 Device

ERROR 2 & 3 Must use template 33

Calibration Procedures

The best method of calibration, if it is possible to do so, is the

LiVE

calibration, as this reads in the sensor signal at

two calibration points and scales the IHH200 automatically. If this is not possible, then the sensitivity figure (in mV/V)

from the sensor calibration certificate can be used to scale the IHH200 by using the

tAbLE

calibration. This may be

the only option available if you are unable to apply a known stimulus to the sensor, which quite often is the case.

LiVE Calibration Procedure

When

CALibrAt

is displayed press

.

LiVE ?

will now be displayed, press

.

You will be prompted

uSE SC ?

. This can be selected if you wish to use the shunt calibration figure from a sensor

calibration certificate (care should be taken that the shunt calibration resistor used originally with the sensor is the

same as is fitted in the IHH). If you wish to use this press

, otherwise press

.

You will then be prompted

APPLY LO.

At this point ensure that the low calibration stimulus is applied to the sensor

and allow to settle of approximately 3 seconds, then press

You then be prompted with

dISP LO.

Press

to enter the display value required with the low stimulus applied to

the sensor. The value can be entered by using the

and

button to select a digit and the

and

buttons to change the digit. When the value has been set press

.

You will then be prompted with

APPLY HI

(unless you chose to

uSE SC ?

, in which case jump to the next stage) At

this point ensure that the high calibration stimulus is applied to the sensor and allow to settle of approximately 3

seconds, then press

.

You then be prompted with

dISP HI

. Press

to enter the display value required with the high stimulus applied to

the sensor. The value can be entered by using the

and

button to select a digit and the

a

nd

buttons to change the digit. When the value has been set press

.

You should now see

donE

displayed. This means the calibration was successful, press

to the IHH200 to

normal operation mode, with the new calibration data stored. If you see

FaiLEd

, then you will need to

repeat the calibration, checking that you have completed the procedure in the correct order, and that the sensor is

connected correctly.

tAbLE Calibration Procedure

When

CALibrAt

is displayed press

.

LiVE ?

will now be displayed, press

.

tAbLE ?

will now be displayed, press

.

You will be prompted with

InPut LO

, press

.

10 Thomas, Irvine, CA 92618

USA T (949)-465-0900 F (949)-465-0905

[email protected]

www.futek.com/support/IHH200

IHH200 Manual 2.5 - Page 21

Template 33 restrictions

ERROR 6

min physical value = >-9999999.0

max physical value = >9999999.0

ERROR 4

value precision case = 1 or 2

ERROR 7

min electrical value > -5.0mV/V

max electrical value < 5.0 mV/V

ERROR 5

bridge type = FULL (2)

ERROR 8

excitation min = > 5.0

excitation max = < 5.0