Sensitivity (+), Loading profile in progress please wait 30 – FUTEK IPM650 Series User Manual

Page 35

Advertising
background image


EM1004-B

- 35 -

10 Thomas, Irvine, CA 92618 USA

Tel: (949) 465-0900

Fax: (949) 465-0905

Toll Free: (800) 23 FUTEK

6.1.3.5 Sensitivity (+)

Scroll to “

SENSITIVITY (+)

” from “

NEW CHANNEL

” using the

▲▼

keys and press ENTER to set the

sensitivity for a new channel. “

SENSITIVITY

” can also be selected from “

EDIT CHANNEL

” to make

modifications on the existing channel except channel one. Sensitivity of the channel can be selected in
mV/V using the

◄►

and the

▲▼

keys.

If the actual sensitivity is not known, but the specific range is known, use the larger number. For example
if the sensitivity of the sensor is between 2 mV/V and 3 mV/V, 3mV/V should be entered. A live
calibration would then be needed to find the actual mV/V. Any time a new sensor sensitivity is set, the
device may do an auto calibration after it loads the new profile.
The following messages will be seen on the screen.
Note: Sensitivity can be defined for bridge sensors and voltage sensors. This option is disabled for
current sensors.

The following message will be seen if the sensor sensitivity is the same as current profile.

OPERATION

WAS EXECUTED

SUCCESSFULLY

(HIT ENTER/BACK)



AUTO CALIBRATION

IN PROGRESS

PLEASE WAIT 15

………



LOADING PROFILE

IN PROGRESS

PLEASE WAIT 30



SELECT SENSOR

mV/V OUTPUT

USING ARROW KEYS

[2.00000]




SENSOR CONFIG

► SENSITIVITY(+)

UNIT SELECTION

SENSOR CAPACITY




Advertising