The equipment, 20 kaman precision products – Kaman KDM-8200 User Manual

Page 20

Advertising
background image


20

Kaman Precision Products

The Equipment

Sensor

The table below lists the variable inductance sensors available for use with the

Specification KDM-8200 family and their standard performance specifications:


SENSOR
MODELS

MEASURING
RANGE
Inch (mm)

TYPICAL
OFFSET
Inch (mm)

ANALOG
VOLTAG
Vdc

DISPLACEMENT
SENSITIVITY
mV/mil (mV/mm)

FULL SCALE
MCD-8000
English (Metric)

.5U/.5U2

.020 (0.5)

.002 (.05)

2.00 (0.5)

100 (1000)

20.00 (_.500)

1U1/1U2

.040 (1.0)

.006 (.13)

.40 (1.0)

10 (1000)

_40.0 (1.000)

2S1

.080 (2.0)

.015 (.38)

.80 (2.0)

10 (1000)

_80.0 (2.000)

2UB1

.080 (2.0)

.015 (.38)

.80 (2.0)

10 (1000)

_80.0 (2.000)

2U2

.100 (2.5)

.025 (.63)

1.00 (2.5)

10 (1000)

100.0 (2.500)

3U1

.120 (3.0)

.020 (.51)

1.20 (3.0)

10 (1000)

120.0 (3.000)

4S1

.160 (4.0)

.020 (.51)

1.60 (4.0)

10 (1000)

160.0 (4.000)

6U1/6U2

.240 (6.0)

.035 (.89)

2.40 (0.6)

10 (1000)

240.0 (_6.00)

15U1/15U2

.600 (15)

.150 (3.81)

.60 (1.5)

1 (100)

_600. (15.00)

30U1/30U2

1.20 (30)

.300 (7.5)

1.20 (3.0)

1 (100)

1200. (30.00)

60U1

2.40 (60)

.600 (15)

2.40 (.60)

1 (100)

2400. (_60.0)

15N(diff)

±.009 (±.25)

.006 (.15)

±9.0 (±9.0)

1000 (36000)

±9.00 (.250)

20N(diff)

±.009 (±.25)

.006 (.15)

±9.0 (±9.0)

1000 (36000)

±9.00 (.250)

1925(HT)

.050 (1.27)

.005 (.127)

2.50 (2.5)

50 (2)

_50.0 (_1.27)

1925M(HT)

.040 (1.0)

.002 (.05)

2.00 (2.0)

50 (2)

_40.0 (_1.00)

1950 (HT)

.150 (3.81)

.010 (.254)

1.50 (1.50)

10 (400)

100.0 (_3.81)

1950M (HT)

.100 (2.54)

.005 (.127)

1.00 (1.00)

10 (400)

100.0 (_2.54)

1975 (HT)

.200 (5.00)

.010 (.254)

2.00 (2.00)

10 (400)

200.0 (_5.00)

1975M (HT)

.100 (2.54)

.005 (.127)

1.00 (1.00)

10 (400)

100.0 (_2.54)


Note: “_xx.x” indicates that the left-most display character (for higher resolution) is not available at the

scale factor selected due to limitations of the meter. The meter has a minimum input requirement of 200
microvolts per count.

The first digit in the sensor number indicates the specified linear measuring range in millimeters. (The

conversion factor from inches to millimeters is .03937.)

The letters “S” and “U” indicate whether the sensors are shielded or unshielded.

Other sensors, including differentially operating ones are available. If your application requires one of these

sensors, a special supplement to this manual will be made available.


Measuring

The specified linear measuring range for an inductive system is directly proportional to

Range

the diameter of that sensor. Changes in oscillator frequency or power to the sensor normally have
no effect on measuring range. For any given sensor diameter, unshielded sensors have a greater
measuring range than shielded sensors. This is because the sensor’s field will “couple” with the
shield, in effect limiting the amount of field available for interacting with the target.

TARGET =

2.5 to 3X diameter of
an unshielded sensor

UNSHIELDED SENSOR

SHIELDED SENSOR

TARGET =

1.5 to 2X diameter of

an shielded sensor