Quadtemp general specifications, Product user guide – MadgeTech QuadTemp Data Logger User Manual

Page 8

Advertising
background image

Product User Guide

4 Internal Channels

Temperature Range -20 ºC to +60 ºC

Temperature Resolution 0.05 ºC

Calibrated Accuracy ±0.5 ºC (0 to +50 ºC)

4 Remote Channels

Remote Channel

Thermocouple Types

J, K, T, E, R, S, B, N

Thermocouple Connection Female subminiature (SMP)

Cold Junction Compensation Automatic, based on internal channel

Maximum Thermocouple

Resistance

1000 Ω, <100 Ω recommended

Thermocouple

Range (ºC)

Resolution

Accuracy

J

-210 to +760

0.1 ºC

±0.5 ºC

K

-270 to +1370

0.1 ºC

±0.5 ºC

T

-270 to +400

0.1 ºC

±0.5 ºC

E

-270 to +980

0.1 ºC

±0.5 ºC

R

-50 to +1760

0.5 ºC

±2.0 º C

S

-50 to +1760

0.5 ºC

±2.0 º C

B

+50 to +1820

0.5 ºC

±2.0 º C

N

-270 to +1300

0.1 ºC

±0.5 ºC

Start Modes

Software programmable immediate start or delay start up
to six months in advance

Real Time Recording

May be used with PC to monitor and record data in real
time

Memory

500,000 readings per channel, channels can be disabled
to increase memory

Reading Rate 4 readings per second up to 1 every 24 hours

Calibration Digital calibration through software

Calibration Date Automatically recorded within device

Battery Type

9 Volt lithium or alkaline battery included,
user replaceable

QuadTemp General Specifications