Time constants at 23°C and 1 m/s air movement:

Humidity < 10 s Temperature < 15 s

Adjustment points (potentiometers)

Humidity

100% limit

Temperature


35%,80%, 10%
Hmax
Tmin, Tmax

Output signals:

Humidity

Temperature

-MP 101A, 111A

-MPI02A,112A

-MPI03A,113A


0...1 VDC = 0...100 %RH

-0.4..+0.6 V DC= -40..+60 °C
0...IVDC = -30..+70 °C
0...IVDC = -40..+60 °C

Load per output

> 1000 Ohms

Supply voltage

-

MP 101A, 102A, 103A

-

MP 111A,112A,113A


4.8...30, VDC
3.6...12VDC

Max. current consumption

8 mA (4.8...30,0 V DC) 7 mA (3.6...12 V DC)

Probe connection

(Standard) 7-pin Tuchel plug

Sensor protection:

Wire filter

Weight

70 g

8.1 Connector wiring

Find the plug connection diagram belonging to your probe
Example:
Code for the standard version with 7- pin Tuchel connector: T7

Connection designations
+

Supply voltage + VDC

GND

Common ground (-), - VDC

RH

Humidity (+)

T

Temperature (+)

0

Cable compensation


View of the connection side of the opposite connector
Standard version with 7-pin Tuchel connector


Example: MPI0IA-T7-W4W













Version with 4-pin Tuchel connector

Example: MPI0IA-T4-W4W