ROTRONIC MP100A User Manual

Page 19

Advertising
background image


Version with Cannon connector

Example: MP101A-CP02-C4-W4W




8.2 Connection cables without connectors

Code CP

PVC cable

Code CG

PUR cable grey

Example:

MP101A -CG-02...


Colour of wire

Supply voltage

+ VDC

green

Relative humidity

% RH (+)

white

Temperature

T (+)

brown

Compensation

0

yellow

Common ground

(-), -VDC /GND

shield

8.3 Cable compensation

This chapter must be observed only if no ROTRONIC display instrument is used for the evaluation.
The integrated cable compensation circuit enables a connection cable up to 100 m long to be used
without influencing the measuring signal.

a) 7-pin Tuchel connector
The connection from the probe to the indicating instrument requires 5
conductors (5-core or 4-core cable + shield). In the 4-core + shield cable
type, the shield is used for the common ground "GND".

For the cable compensation to function, the compensation "0" must be
connected with the common ground "GND" on the load-side socket
(display, controller, transducer).





If the five-core cable is prolonged with a five-core cable but concluding
with a 4-pin connector, one can connect to this the compensation "0" with
the common ground "GND". In this case the cable compensation acts in
the five-core cable up to the 4-pin connector.

b) 4-pin Cannon connector with cable

The cable compensation is effective up to the connector.



d) 4-pin Tuchel connector

The cable compensation is not effective in this case.