5 set input 1 or 2 for a digital or analog probe, 5 data logging function, 1 logging data – ROTRONIC HF8 User Manual

Page 26: Data logging function, E-m-hf8-v2_10

Advertising
background image

E-M-HF8-V2_10

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

HygroFlex HF8 Humidity Temperature
Transmitter: User Guide

Instruction Manual

Document Type

Page 26 of 35

Document title

© 2012; Rotronic AG E-M-HF8-V2_10

7.4.4

Select which probe and/or parameters are shown on the display

Press the MENU key and select Device Settings > Display Settings > Mode. Press ENTER to activate the
Mode menu item, use the [+] or [-] key to select the display mode. Press ENTER to confirm and press MENU
to exit.

Depending on the display mode, use the [+] or [-] key to change the probe and/or parameter being displayed.

NOTE: The calculated parameter (HygroClip 2 probe only) is shown only if enabled for the probe input that is
selected (MENU > Input 1 or Input 2 > Calc).

7.4.5

Set input 1 or 2 for a digital or analog probe

Press the MENU key and select Device Settings > Input 1 or Input 2 > Pbe Type. Press ENTER to activate the
Pbe Type menu item, use the [+] or [-] key to change the probe type. Press ENTER to confirm and press
MENU to exit.

When using an analog probe, be sure to define both the voltage signal range and the measuring range of the
probe. HW4 is required to define the unit of measurement of an analog probe.

The HW4 software can also be used to change the probe type for each input.

7.5 Data logging function


The HF8 can log up to 10,000 pairs of relative humidity and temperature values provided by a single
HygroClip 2 probe or up to 10,000 data values provided by a single 1-channel analog probe. Both probe
inputs can be logged at the same time and in that case the recording capacity per probe is cut in half. Each
record is automatically date and time stamped. The calculated parameter cannot be recorded.

7.5.1

Logging data

The data logging settings apply to both probe inputs

Data logging starts and ends simultaneously for both probe inputs

The log data function settings cannot be changed as long as data logging is active

Configure the data logging function and start recording data:

o

Press the MENU key and select “Data Logging”. Press ENTER to activate the Data Logging menu.

o

Use the + or - key to select Settings. Press ENTER to confirm and open the Settings sub-menu. Use
the + or - key to select a menu item and press ENTER to confirm:

o

Select Interval (log interval). Press ENTER to activate the Interval menu item and use the + or - key
to change the log interval. Press ENTER after each change to confirm and move the cursor to the
right. When done, press ENTER to confirm and exit.

o

Use the + or - key to select the Mode menu item. Press ENTER to activate the Mode menu item and
use the + or - key to change the logging mode:
- Start-Stop. the recording will stop when the memory is full
- Loop: when the memory is full the oldest record will be dumped to make room for the next record
When done, press ENTER to confirm and exit.

o

Use the + or - key to select each probe to be logged. Press ENTER to activate the Probe 1 or Probe
2 menu item and use the + or - key to enable data logging. Press ENTER to confirm and exit.

o

Press the MENU key and use the - to select Start Recording

o

Press the ENTER key twice to start recording data

o

The HF8 automatically exits the data logging function and a diskette symbol appears at the bottom
left of the display for each probe being recorded