5.0 MODEL 227 PERFORMANCE SPECIFICATIONS

Accuracy RSS

*

(at constant temperature.)

±1.0% Reading

±0.25% FS

Non-Linearity, BFSL

±0.15% FS

Hysteresis

0.20% FS

Non-Repeatability

0.02% FS

*RSS of Non-Linearity, Non-Repeatability and Hysteresis.

Thermal Effects
Compensated Range

°F(°C)

+15 to +150 (-9 to +65)

Zero Shift %FS/100

°F (50°C)

±2.0 (±1.8)

Span Shift %FS/100

°F (50°C)

±2.0 (±1.8)

Environmental
Temperature
Operating**

°F(°C)

-40 to +185 (-40 to 85)

Storage

°F(°C)

-40 to +185 (-40 to 85)

Current unit ordered with Option N1:
Operating**

°F (°C)

-22 to +176 (-30 to +80)

Storage

°F (°C)

-22 to +176 (-30 to +80)

**Operating temperature limits of the electronics only. Pressure media temperatures may be considerably higher or lower.

6.0 RETURNING PRODUCTS FOR REPAIR

Please contact a Setra application engineer (800-257-3872, 978-263-1400) before
returning unit for repair to review information relative to your application. Many
times only minor field adjustments may be necessary. When returning a product to
Setra, the material should be carefully packaged, accompanied by the “Service/
Repair/Calibration Order Return” form on page 7 and shipped to :

Setra Systems, Inc.
159 Swanson Road
Boxborough, MA 01719-1304
Attn: Repair Department

Note: Return order form is also available on Setra’s web site @ www.setra.com.

To assure prompt handling, please supply the following information and include it
inside the package or returned material (Return Order Form provided on page 7):

1.

Name and phone number of person to contact.

2.

Shipping and billing instructions.

3.

Full description of the malfunction.

4.

Identify any hazardous material used with product.

Notes:
Please remove any pressure fittings and plumbing that you have installed and enclose any
required mating electrical connectors and wiring diagrams. Allow approximately 3 weeks after
receipt at Setra for the repair and return of the unit. Non-warranty repairs will not be made
without customer approval and a purchase order to cover repair charges.

Calibration Services

Setra maintains a complete calibration facility that is traceable to the National
Institute of Standards & Technology (NIST). If you would like to recalibrate or recertify
your Setra pressure transducers or transmitters, please call our Repair Department at
800-257-3872 (978-263-1400) for scheduling.