0 operation, 1 analogue output probes, 2 sdi12 operation – Setra System Model 595 User Manual

Page 6

Advertising
background image

It is recommended to firmly secure the Kevlar core of the cable, for example, by
knotting and securing under a fixing screw.

7.0 Operation

Having installed the transducer as instructed it is ready for use. Before applying
power, check that the correct polarity and excitation levels are being applied. See
Electrical Installation (6.2).

7.1 Analogue Output Probes

Analogue output devices will simply provide a 4 mA output for 0 level and a 20
mA output for the full scale level indicated on the product label. The analogue
output can be adjusted via the digital SDI12 interface with a suitable accessory to
include site offsets or provide full scale, 20 mA, output at a different level.

7.2 SDI12 Operation

The SDI12 interface conforms with the SDI12 V1.3 specification. Refer to the SDI12
host web-site for a full specification, http://www.sdi-12.org.

In addition to the standard SDI12 command set, the 595 supports a number of
extended commands, providing increased levels of functionality. All sample
commands provided are shown with address “a” , to use the command substitute
the actual address of the unit. The default address when manufactured is “0”.

7.2.1 Setting up the Measured Variable

The transducer can provide either an output proportional to the applied pressure
or a true level compensated for relative density of water. The preferred measured
value must be set by the XP command as detailed in Table 3.

Table 3

Setting the Measured Variable

Command

Response

Comment

aXP0!

a0<CR><LF>

Set true level measure, compen-
sated for relative density of water
over temperature and local gravity.

aXP1!

a1<CR><LF>

Set pressure measure, no additional
compensation.

7.2.2 Making a Level Measurement

The corrected level measurement is calculated in the following manner.

Level = mH

2

O pressure at 4

°C

*

1

*

9.80665

density of water

local gravity

where:

Density of water = -6.017777e-6t

2

= 0.0000408t+0.999841

where t - temperature in

°C.

Table 12

Calibration of the mA Output

Command

Response

Comment

aXM4

a4.00000<CR><LF> Sets the mA output to 4 mA.

Measure the current and record the actual
value.

aXM20

a20.0000<CR><LF> Sets the mA output to 20.0 mA.

Measure the current and record the actual
value.

Step 1:

Apply Zero Scale Value
Set Point Value 4mA, actual reading 4.005 mA.

Step 2:

Apply Full Scale Value
Set Point Value 20mA, actual reading 19.995 mA.

Step 3:

Calculate the Errors and Adjustment Values
The value to write to the offset calibration adjustment is calcu-
lated by subtracting the measured value of 4mA from the 4mA
set point as shown:
Set point zero = 4.000
Measured zero = 4.005
Offset Value = 4.000-4.005 = -0.005 mA

Step 4:

Write the Calibration Adjustment Values
The value to write to the gain calibration adjustment is calculated
by dividing the set point span by the measure span as shown:
Set point span = (20-4) = 16 mA
Measured span = (19.995-4.005) = 15.99 mA
Gain value = 16/15.99 = 1.0006254

Table 13 shows the command required to write the calibration adjustment to

the 595:

Table 13

Writing the mA Output Calibration

Command

Response

Comment

AXZ3,-0.005!

a-0.00500<CR><LF>

Writes the offset of -5

µA to the mA

output zero.

AXG3,1.0006254! A1.006254<CR><LF>

Writes the gain value t the mA
output.

Step 5:

Apply Zero Scale Value
Set Point Value 4mA, actual reading 4.001mA.

Advertising