OPERATING INSTRUCTIONS

MODEL 141 ACCELEROMETER

GENERAL INFORMATION

Your Setra accelerometer has been carefully calibrated before shipment to you, and it should be
handled with the same care given any precision instrument. Setra accelerometers are rugged devices,
built to take considerable overrange of “G” loads.

Please refer to the Model 141 calibration data before proceeding further.

ELECTRICAL CONNECTIONS

Cable Lead

Function

White

positive excitation

Black

negative excitation

Yellow

positive output

Brown

negative output

Red

calibration signal

Shield

case

CAUTION:

BE ABSOULUTELY SURE OF EXCITATION POWER POLARITY BEFORE
APPLYING POWER, REVERSED EXCITATION WILL BURN OUT THE CIRCUIT
INSTANTLY.

MODEL 141A MUST NOT BE OPERATED WITH EXCITATION EXCEEDING 15 VDC.

Model 141A accelerometer excitation voltage 5 VDC to 15 VDC. Keep excitation power between 5 VDC
min. and 15 VDC maximum. High voltage may burn out the circuit. Installation of a protective Zener
diode such as a IN4746A across the excitation power input leads is a good practice, taking care to

0.225 (max)
5.715

.

Center of Seismic Mass
Also Center of Gravity of
Accelerometer (less cable)

+G

0.875 (max)
22.225

Cable Length 2 feet
Unless Otherwise Specified

0.328
8.331

4 Mounting Holes
0.125/0.3175
Dia.

±

0.002/0.0508

Within

±

0.005/

±

0.127

of True Position

0.656

16.652

Brown - Output
Black - Excitation

Yellow
+ Output

Red
Calibration

White
+ Excitation

Cable Clamp Screws
0-80 x 5/64 Pan head

Recommended Mounting Screws
No. 4-40 U.N.C. Class II
Stainless Steel Socket Head
Cap Screws 0.5/12.7 Long Minimum

0.880 SQ (max)
22.352

0.689
(max.)

0.332/ 8.432

±

0.004/0.1016

0.328
8.331

0.656

16.652