3000RS PRECISION RESISTANCE STANDARD
OPERATION MANUAL

Transmille Ltd.

Page 6

Label Reference

Information

Model Number

Model number reference for product

Serial Number

Unique serial number for product

Operating Notes

This section details operational and environmental considerations for the 3000RS

precision resistance standard. Follow these instructions when operating or storing

the resistance standard.

Connecting to the 3000RS

CAUTION

Do not apply voltages above the maximum ratings detailed below. Excessive voltage can

cause permanent damage and affect the nominal resistance values of the reference.

VALUE

MAXIMUM VOLTAGE RATING

1 Ohm

0.5V

10 Ohms

1V

100 Ohms

10V

1kOhm 10V

10kOhms 50V

100kOhms 100V

1MOhm 100V

10MOhm 300V