3000RS PRECISION RESISTANCE STANDARD
OPERATION MANUAL

Transmille Ltd.

Page 7

The specifications for connections with CURRENT and VOLTAGE (SENSE) require four wire

connections as standard. Use these terminal posts are detailed in the section below.

For the case to act as a shield the EARTH terminal must be connected to earth ground or

instrument guard at some point in a calibration configuration. When connecting two or more

instruments with guards, the use of a single common earth point avoids earth loops.

Connection Recommendations

For optimum results use two sets of low thermal test leads which are designed to reduce

errors caused by thermal EMFs.

Fig. 1 : 4-Wire Connection Example