Transmille EA3023 User Manual

Page 14

Advertising
background image

EA3023 Power Supply Calibration Adapter Specifications

General Specifications
Adaptor connection

Connects to male 'D' type adapter interface on 3000 Series front panel

Indicators

Incorporates 1x Power indicator LED : 1 x Overload indicator LED

Connections

1 x 9 way male 'D' type connector
2x Voltage input 4mm terminals (Red & Black)
2x Current load 4mm terminals (Red & Black)

Connection to Calibrator Via supplied 9 Way male to female serial lead (straight through connection)

DC Voltage Measurement Accuracy
Range

Resolution

90 Day Rel.

180 Day Rel.

1 Year Rel.

2 Year Rel.

%

±

mV

%

±

mV

%

±

mV

%

±

mV

0 to 63 Volts

0.01V

0.016 ±

20

0.016 ±

20

0.02 ±

20

0.028 ±

20

DC Current Load Accuracy
Range

Resolution

90 Day Rel.

180 Day Rel.

1 Year Rel.

2 Year Rel.

%

±

mA

%

±

mA

%

±

mA

%

±

mA

0 to 3 Amps

0.01A

0.040 ±

1.0

0.045 ±

1.0 0.05 ±

1

0.070 ±

1.0

Specifications apply between 17°C and 27°C.

Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Due to continuous development specifications may be subject to change.

2000 Series Options Extended Specifications : Power Supply Calibration Option