2090 OPERATION MANUAL

TRANSMILLE LTD.

PAGE 6

Remote Operation


To enable remote operation, the 2090 must be connected to a PC using the built in RS232
interface (located on the rear of the instrument).

Connection to RS232 COM port

Connection to USB port (using RS232 to USB converter)

Once connected, using the PC the resistance value should be sent to the 2090 to set the
output. The command MUST be terminated with a carriage return (ASCII character 13),
for example :

COMMAND DESCRIPTION DISPLAY

0.2

<CR>

sets the display to

0.2 Ohms

00000.2 Ohms

100.0

<CR>

sets the display to

100 Ohms

00100.0 Ohms

1000.0

<CR>

sets the display to

1000 Ohms

01000.0 Ohms


Where <CR> is a carriage return (ASCII 13)

The value must include a decimal point, i.e. 100 Ohms should be sent as 100.0

Note : The front panel controls are not locked out when in remote operation mode, to enable
manual adjustment of any setting.