P50 PRESSURE COMPENSATION

When P50 and "PRES" (Fig. 29) are displayed

on the LCD, press to set pressure

compensation value for CO² measurement.

The current set will flash on LCD. (Fig. 30)

The barometric pressure unit is kPa.

Press to increase or to decrease

barometric pressure.

-OR- Press to escape and return to P50.

NOTE: ONLY ENTER THIS MODE IF CURRENT

BAROMETER PRESSURE IS KNOWN. MULTIPLY IN.
Hg BY 3.39 TO OBTAIN kPa VALUE

.

P60: REAL TIME CLOCK

Enter P60 for setting the real time clock of

this meter. Press to enter into P61 to set

the time format as 12 hour or 24 hour. The

current set will flash on LCD. Press

or to change the format and press

to confirm and continue to set the date

and time. (Fig 32)

-OR- Press to escape and return to P60.

NOTE: DATE FORMAT IS YY:MM:DD.
The date will show on the lower display with

year digits blinking. Press to increase or

to decrease digits. Press to

confirm and continue to next setting. Repeat

above steps to complete the year/month/day/

hour/minute/second setting. (Fig. 32)