Maintenance, Specifications, Inf195 – UEi Test Instruments INF195 User Manual

Page 2: Limited warranty, Infrared thermometer

Advertising
background image

WARNING!

After measurement of high temperature, the probe

may be hot for a while.

Additional Features

Changing Scales

To change scales between degrees Fahrenheit

and degrees Celsius, press the button on the panel,

(marked °F / °C), while the display is active. Even if the

trigger is released and the display is in its 60-second

automatic hold, you can convert the reading between

scales. Each time you press the scale button, the

60-second hold will reset. The INF195 will default to

the scale last used the next time it is turned on.

Using the Laser Sight

The INF195 is classified as a “Laser Product”, and is

regulated by the FDA.

CAUTION!

Using controls, making adjustments, or performing

procedures in any manner other than that specified

herein may result in hazardous radiation exposure.

LCD Error Messages

The thermometer incorporates visual diagnostic

messages as follows:

: “Hi” or “Lo” is displayed when the

temperature being measured is

outside the settings of HAL and LAL.

: “Er2” is displayed when the

thermometer is exposed to rapid

changes in the ambient temperature.

: “Er3” is displayed when the ambient

temperature exceeds the operation

range of 32˚ to 122˚F (0˚ to 50˚C). The

thermometer should be allowed

plenty of time (minimum 30 minutes)

to stabilize to the working/room

temperature.

: For all other error messages it is

necessary to reset the thermometer.

To reset the thermometer, turn the instrument off, remove the

battery and wait for a minimum of one minute. Reinsert the

battery and turn it on. If the error message remains please

contact UEi service department for further assistance.

Battery Indicators

The thermometer incorporates visual low battery

indications.

: “Battery OK” measurements

are possible

: “Battery Low” battery needs to be

replaced, measurements are

still possible

: “Battery Exhausted” measurements

are not possible

Maintenance

Periodic Service

WARNING!

Repair and service of this instrument is to be performed

by qualified personnel only. Improper repair or service

could result in physical degradation of the instrument.

This could alter the protection from electrical shock and

personal injury this instrument provides to the operator.

Cleaning
Periodically clean your instrument’s case using a damp

cloth. DO NOT use abrasive, flammable liquids, cleaning

solvents, or strong detergents as they may damage the

finish, impair safety, or affect the reliability of the struc-

tural components.

Battery Replacement
Always use a fresh replacement battery of the specified

size and type. Immediately remove the old or weak bat-

tery from the meter and dispose of it in accordance with

your local disposal regulations. Old or defective batteries

can leak chemicals that corrode electronic circuits.

To replace the battery:

1. Slide the battery cover straight down then away

from the INF195.

2. Replace with fresh “AAA” (2x) batteries.

3. Replace cover.

Lens Care
The sensor lens is the most delicate part of the ther-

mometer. The lens should be kept clean at all times. Care

should be taken when cleaning the lens using only a

soft cloth or cotton swab with water or medical alcohol.

Allowing the lens to fully dry before using the thermom-

eter. Do not submerge any part of the thermometer in

liquids.

WARNING!

Under NO circumstance should you expose batteries

to extreme heat or fire as they may explode and cause

injury.

NOTE:

This instrument contains no user serviceable

parts. If service becomes necessary, call UEi and ask for

the service department. See the warranty section of this

manual for additional details.

INF195

Infrared Thermometer

Limited Warranty

The INF195 is warranted to be free from defects in materials and work-

manship for a period of three years from the date of purchase.

If within the warranty period your instrument should become

inoperative from such defects, the unit will be repaired or replaced

at UEi’s option. This warranty covers normal use and does not cover

damage which occurs in shipment or failure which results from

alteration, tampering, accident, misuse, abuse, neglect or improper

maintenance. Batteries and consequential damage resulting from

failed batteries are not covered by warranty.

Any implied warranties, including but not limited to implied warranties

of merchantability and fitness for a particular purpose, are limited to

the express warranty. UEi shall not be liable for loss of use of the

instrument or other incidental or consequential damages, expenses, or

economic loss, or for any claim or claims for such damage, expenses or

economic loss. A purchase receipt or other proof of original purchase

date will be required before warranty repairs will be rendered.

Instruments out of warranty will be repaired (when repairable) for a

service charge. Return the unit postage paid and insured to;

1-800-547-5740 • FAX: (503) 643-6322

Service: (800) 308-7709

www.ueitest.com • Email: [email protected]

This warranty gives you specific legal rights. You may also

have other rights which vary from state to state.

Specifications

EMC/RFI

Readings may be affected if the unit is operated within

radio frequency electromagnetic field strength of

approximately 3 volts per meter, but the performance

of the instrument will not be permanently affected.

IR Thermometers
Infrared Thermometers measure the amount of thermal

radiation emitted from an objects surface. To take a

measurement, the object must be within the field of

view of the infrared thermometer, or the reading may

include objects surrounding your target. This target size

is determined in the distance to spot ratio stated on the

instruments specifications. At a ratio of 9:1, a distance

from your target of nine feet will give you a circle with

one foot diameter. At 20:1 ratio gives the same one foot

target now at 20 feet, so you can see that higher is

better, but typically more expensive due to the cost of

the optics involved.

The other common question is regarding emissivity. In

simple terms emissivity is a percentage of energy

emitted from a surface compared to the energy emitted

from a black body source. If a surface emits one half

(1/2 or 0.5) the amount of energy at a given temperature

and wavelength as a black body, it is said to have an

emissivity of 0.5.

Surfaces closest to the balck body level would be flat

black, and those farthest from it would be mirror or

chrome surfaces. With a set emissivity level some

surfaces may measure lower than actual, because their

surface is emitting less thermal radiation at a given

temperature. Although you may have a fixed emissivity

of 0.95, most items measured will provide a reasonably

accurate result. Having a fully adjustable emissivity will

give you the ability to fine tune your instrument to a

specific application.

To determine the emissivity of the surface being tested,

paint a portion with flat black paint, and then measure

the temperature of the painted area and compare it

to the temperature of the standard surface. Use the

following to determine the emissivity to use.

Temperature (Standard area) divided by temperature

(painted area). As an example 61° in the standard area

and 68° painted is 61/68 or 0.89

INF195-MAN 11/11

Copyright © 2011 UEi

PLEASE

RECYCLE

D:S=1

2:1

6.67@80”

3@36”

2@24”

INF1

95

Infrared

Thermomete

r

Item

Non-contact

Thermocouple Probe

Infrared Scan

Scan

Measurement range -76˚ to 1022˚F

-83.2˚ to +1999˚F

(-60˚ to +550˚C)

(-64˚ to ≠1400˚C)

Operating range

32˚ to 122˚F (0˚ to +50˚C)

Accuracy

±2% RDG or 2˚C (4˚F)

±1% of reading or 1.8˚F

(Tobj= 0~550˚C)

whichever is greater

(1.0˚C) whichever is

greater (Test under

(Tobj= -60 ~ 0˚C)

((±2˚C (4˚F)

Tamb=73±6˚F)

+ 0.05˚/per degree)

Resolution

0.1˚F / 0.1˚C

(-9.9-199.9˚F)
Response time (90%) 1 second
Distance: Spot

12 : 1

Battery life

Standard “AAA” (x2), 140 hours

Dimensions

6.9” x 1.5” x 2.8” (175.2 x 39.0 x 71.9mm)

Weight

6.7 oz

NOTE: Under the electromagnetic field of 2V/m from 200 to 600 MHz, the maxi-
mum error is 18˚F (10˚C).

INF195

“AAA” (x2)

Perform this first

Verify that the meter

Press to activate the following

is in one of these modes

Press and release

In MIN, MAX, DIF,

LOCK - Press the “ “ push-button for LOCK

the trigger to activate AVG mode

mode ON/OFF - This locks the IR measurement

display first

for up to 60 minutes

SCALE - Press “ “ push-button to select ˚C or ˚F

Press and hold

In all modes except PRB

BACKLIGHT/WORKLIGHT - Press the “ “

the trigger to use

(Probe)

push-button for backlight function ON/OFF

these functions

LASER - Press the “ “ push-button for laser

function ON/OFF

Advertising