Agilent Technologies E1364A User Manual

Page 28

Advertising
background image

Normally Closed to

Common

1. Hardware Connections

2. Equipment Setup

Set DMM to: 2-wire ohms, 1 G

range.

Send

*RST

to the switch.

Send the

CLOS (@100:115)

command to the switch to open all NC

contacts.

3. DC Isolation Reading

Trigger the DMM and record the reading.

Figure 2-5. DC Isolation Test (NC Channels to Common)

28 Verification Tests

Chapter 2

Advertising