Example: dc isolation test – Agilent Technologies E1364A User Manual

Page 50

Advertising
background image

Example: DC

Isolation Test

This example performs the DC Isolation Test for Normally Open and
Normally Closed channels to both Common and chassis (earth ground).

/* DC Isolation Test E1364A */

#include <stdio.h>
#include <sicl.h>

#define ADDR "hpib7,9,15"

/* Address of device */

#define DMM "hpib7,22"

void main (void)
{
INST id, dm;

/* Define id and dm as an instrument */

char reading[256] = {0};

/* Result variable */

int channel, i;

#if defined(__BORLANDC__) && !defined(__WIN32__)
_InitEasyWin();
#endif

ionerror(I_ERROR_EXIT);

id = iopen (ADDR);

/* Open instrument session */

dm = iopen (DMM);

iprintf (dm, "PRESET NORM;TRIG HOLD\n");
iprintf (dm, "FUNC OHM;RANGE 1E9\n");

printf ("\n\nConnect DMM HI and LO to E1364A NO and C lines");
getchar ();
iprintf (id, "*RST\n");
ipromptf (id, "*OPC?\n", "%t", reading);
ipromptf (dm, "TRIG SGL\n", "%t", reading);
printf ("\nDC Isolation -- Normally Open Channels to Common");
printf ("\n R = %s", reading);

printf ("\n\nConnect DMM HI and LO to E1364A NC and C lines");
getchar ();
iprintf (id, "CLOS (@100:115)\n");
ipromptf (id, "*OPC?\n", "%t", reading);
ipromptf (dm, "TRIG:SGL\n", "%t", reading);
printf ("\nDC Isolation -- Normally Closed Channels to Common");
printf ("\n R = %s", reading);

50 Example C Programs