0 operational check and oxygen concentration test, 1 description of operation, 2 operation check – AirSep MN134-1 B User Manual

Page 10: 3 alarm system

Advertising
background image

MN134-1 rev B

02/14

5

2.0 Operational Check and Oxygen Concentration Test

2.1 Description of Operation

Air is drawn into the NewLife Intensity 10 Oxygen Concentrator through an external air
intake gross particulate filter. Before this air enters the compressor, it passes through the
unit’s suction resonator which, quiets the compressor’s suction sound. Pressurized air then
exits the compressor and passes through a heat exchanger. The heat exchanger reduces the
temperature of the compressed air. Next, a two-way solenoid feed valve directs the air into
one of two sieve beds that contain molecular sieve. The unique property of molecular
sieve enables it to physically attract (adsorb) nitrogen when air passes through this
material, thus producing high concentrated oxygen.

There are two sieve beds: while one produces high concentration oxygen, the other is
purged of the nitrogen it adsorbed (collected) during this pressure swing adsorption (PSA)
cycle. Each adsorber produces oxygen and delivers it to the product tank. Oxygen exits
the product tank through a pressure regulator, flow control valve, and flowmeter. The
NewLife Intensity 10 unit delivers up to 95.5% oxygen at flow rates from 2-10 lpm.

2.2 Operation Check

AirSep tests every NewLife Intensity 10 Oxygen Concentrator thoroughly after
manufacture. You must perform the following test to ensure that no damage occurred in
shipping or handling.

Do not use extension cords with this unit or connect too many plugs
into the same electrical outlet. The use of extension cords could
adversely affect the performance of the device. Too many plugs into
one outlet can result in an overload to the electrical panel, causing the
breaker/fuse to activate or fire if the breaker or fuse fails to operate.

1.

Open and inspect all cartons (that contain units) upon delivery. Unpack the unit
and remove it from the carton. Inspect the unit itself for damage. If the exterior of
a unit’s carton is damaged, or the unit itself is damaged, note it on the freight bill
signed by the driver.

2.

Connect unit to power, and set the

I

/0 power switch to the “

I

” position. Check to

see that the following occurs:

a. A continuous alarm sounds for approximately five seconds. See the

troubleshooting chart in Section 6.3 of this manual if the unit’s
alarm does anything other than sound for five seconds.

b. The compressor runs and flow is indicated in flowmeter.
c. OPTIONAL: The Oxygen Monitor’s amber light remains on until

the oxygen concentration reaches 85% ±3% (approximately two
minutes). Perform an oxygen concentration test, as described in
Section 2.4.

2.3 Alarm System

The NewLife Intensity 10 Oxygen Concentrator is equipped with a battery-powered alarm
system, which sounds a continuous and loud alarm when a power failure occurs. It sounds
an intermittent alarm if the high or low pressure indicators are activated or if the optional

Advertising