0 patient instructions, 1 instructions, 2 routine maintenance by the patient – AirSep MN134-1 B User Manual

Page 12: 1 cleaning the air intake gross particle filter, 2 checking the alarm system battery

Advertising
background image

MN134-1 rev B

02/14

7

3.0 Patient Instructions

3.1 Instructions

It is important that patients thoroughly understand how to operate the AirSep NewLife
Intensity 10 unit. This enables proper treatment, as prescribed by a licensed health care
provider/physician. If patients experience any discomfort or the unit alarms, they must
notify their licensed health care provider/physician immediately.

You, as the Equipment Provider, are responsible to see that each patient receives the
Patient Manual. Explain each step in the operation of the unit to the patient in reference to
this manual.

This device supplies high-concentration oxygen that promotes rapid
burning. Do not allow smoking or open flames within the same room of
(1) this device, or (2) any oxygen-carrying accessory. Failure to observe
this warning can result in severe fire, property damage, and /or cause
physical injury or death.

3.2 Routine Maintenance by the Patient

To ensure accurate output and efficient operation of the unit, the patient must perform two
simple routine maintenance tasks:

§

Clean the air intake gross particle filter

§

Check the alarm system battery

3.2.1 Cleaning the Air Intake Gross Particle Filter

The patient must clean this filter weekly, as described below. The
filter may require daily cleaning if the NewLife Intensity 10 unit
operates in a harsh environment such as a house heated by wood,
kerosene, or oil, or one with excessive cigarette smoke.

1. Remove the dirty air intake gross particle filter from the back of the NewLife

Intensity 10 unit, and install a clean filter, as described above.

2. Wash the dirty filter in warm soapy water, and rinse.
3. Use a soft absorbent towel to remove excess water.
4. Keep the clean, dry filter readily available for next use.

3.2.2 Checking the Alarm System Battery

The alarm system battery is tested each time the

l

/0 power switch is set to the “

l

position. A continuous alarm sounds for approximately five seconds to indicate a
good battery. If the alarm does anything other than sound for five seconds, instruct
the patient to call the Equipment Provider immediately.

Advertising