MN187-1 Rev -

Page 7 of 8

4.

Connect the other end of the MI347 cable to J9 on the Focus circuit board. The
black wire goes to the top pin. (Fig. 14)

Fig. 14

5.

Put the unit in test mode as follows:

Note: Communication is only possible when the Focus unit is in test mode.

5.1

.

Connect the Focus Oxygen Concentrator to an AirSep-supplied AC power

supply (AirSep Part No. PW019-1).








5.2

.

To activate the unit’s test mode:

5.2.1 Turn the unit on.

5.2.2 After the 10 second start up time, turn the unit off & on consecutively

3 times (within 4 seconds).


5.2.3 The unit will then start to auto pulse and is now in test mode.

5.3.

Verify Focus is pulsing by observing the intermittent green LED.
When the unit pulses, you should be able to hear the pulse of oxygen and
see the LED turn off momentarily.

If unit is not pulsing, turn unit off and repeat steps 5.2.1 and 5.2.2.

6.

The Focus Data Interface program screen will display hours in the Unit Run Time
box and the Compressor Run Time box within two minutes. (Fig. 17)

Focus units
Power Inlet

Fig. 16

Focus units
power
connector

Fig. 15