Appendix b model 2530 1ghz calibrator – Boonton 4530 Peak Power Meter User Manual User Manual
Page 169
Boonton Electronics
Appendix B
4530 Series RF Power Meter
2530 1GHz Calibrator
B-1
Appendix B
Model 2530 1GHz Calibrator
MODEL 2530 1 GHz CALIBRATOR
The Model 2530 operates as an optional accessory of the Model 4531 and 4532 RF Power Meters to provide calibration
and test signals at an output frequency of 1.024 GHz. The Model 2530 is controlled by the power meter through its
manual and remote programming systems and GPIB commands are defined for it. The 1.024 GHz calibrator is necessary
for sensors that cannot be calibrated at 50 MHz because of a higher low-frequency cutoff. It is also closer to commonly
used frequencies for telecom applications, and can reduce measurement uncertainty by requiring less interpolation of
sensor frequency calfactors. The Model 2530 calibrator must be conected to the 4530 powermeter using the supplied
RJ-11 control cable as shown in the diagram on the following page.
Accessories
Supplied accessories:
1 - NEMA type power cable
1 - RJ-11 Control Cable
1 - Fuse kit
1 - 2530 Operators Instruction Manual
Other accessories:
Rack Mounting Kit
Specifications
RF Section
Operating Modes:
CW, Internal or external pulse
Output Frequency:
1.024 GHz ± 0.005%
Level Range:
-60.0 to +20.0 dBm
Resolution:
0.1 dB
Source SWR (Refl. Coef.):
1.20 (0.091) maximum, CW mode
Accuracy (NIST traceable):
+20 to -40 dBm
Absolute:
± 0.065 dB (1.5%) at 0 dBm, CW mode
Linearity:
± 0.03 dB per 5 dB from 0 dBm, CW mode
Harmonics:
-35 dB minimum to 40 GHz
Spurious:
-60 dB minimum
Pulse Generator
Internal Pulse Period:
100 ms, 1 ms or 10 ms
Internal Pulse Duty Cycle:
10% to 90% in 10% increments
Pulse polarity:
+ or – for Internal and External pulses
RF Connector:
Type N, front or optional rear panel location
External Pulse Modulation Input:
Type BNC, rear panel, TTL compatible
Controller connector:
RJ-11
Controller:
Boonton Electronics Model 4531/4532 RF Power Meter