5.4.3 Operating the transmitter via the serial bus

5.4.3.1

Setting the serial interface

SMODE xxxx<cr>

xxxx

=

STOP, RUN or POLL

In STOP mode: measurements output only by command, all commands can be

used

In RUN mode:

outputting automatically, only command S can be used

In POLL mode: measurements output only with command SEND. When in

POLL mode, the output mode is changed as follows:

OPEN aa <cr>

SMODE xxxx<cr>

aa

=

address of the transmitter

xxxx

=

STOP, RUN or POLL

The OPEN command sets the bus temporarily in STOP mode so that the
SMODE command can be given.

Example:

>SMODE STOP <cr>

setting STOP mode

Serial mode : STOP