M210300en

User's Guide

59

Disconnect the Pt 100 simulator and reconnect the Pt 100 wires to solder lugs
TP1, TP2 and TP3.

The correct connections according to wire colours are:

TP1

TP2

TP3

blue

white

red

If a Pt 100 simulator is not available, the adjustment can be made with two
resistors of 84

and 154

with precisely known resistance. Measure

resistors with a resistance meter. Look up the corresponding temperature value
from a Pt 100 conversion table or calculate it according to the following
equation:

T = D0 +R x {D1 + R x [D2 + R x (D3 + R x D4)]}

where

D0

=

-243.5673014

D1

=

2.278542701

D2

=

0.002050681

D3

=

-6.15025E-06

D4

=

1.34949E-08

7.5

Temperature channel adjustment with Pt 100 simulators

(composite sensor)

Switch the power off and disconnect the wires to the Pt 100 sensor from solder
lugs TP5, TP6 and TP7.

X

7

7

X

8

8

TP6 TP7

TP5

Figure 7.3

Location of solder lugs TP5, TP6 and TP7 and connector

X88

Connect a Pt 100 simulator to connector X88 and set it at the lowest tempera-
ture to be calibrated.