Show analog output status – Vaisala HMT130 User Manual

Page 35

Advertising
background image

Chapter 4 ________________________________________________________ Serial Line Operation

VAISALA ________________________________________________________________________ 33

Show Analog Output Status

The AOUT command shows the following information:

- Analog out mode is the output mode for the channel, for example

0 ... 10 V.

- Error level is the level that the analog output is set to in case of

transmitter error.

- Status indicates the current status of the channel. The statuses are

- ON: Normal measurement operation.
- OFF: No quantity selected for the channel using the ASEL

command. See section Set/Show Analog Output Quantity and

Scaling on page 42.

- ERROR: Transmitter error, channel set to error level.
- TEST: Analog channel being tested using the ATEST command.

See section Test Analog Outputs on page 44.

AOUT<cr>

Example:

>aout

*** ANALOG OUTPUT 1 ***

Ch1 Analog out mode : 0_10V

Ch1 Error level : 10.8

Ch1 Quantity : RH

Ch1 Status : ON

RH lo : 0

RH hi : 100

RH : 36.64 %

Ch1 Voltage : 3.66 V

*** ANALOG OUTPUT 2 ***

Ch2 Analog out mode : 0_10V

Ch2 Error level : 10.8

Ch2 Quantity : T

Ch2 Status : ON

T lo : -60

T hi : 100

T : 23.85 'C

Ch2 Voltage : 5.24 V

>

Advertising