User's Guide _______________________________________________________________________

80 ___________________________________________________________________ M211280EN-B


Table 11

Operating Environment Specifications

Property

Description/Value

Operating temperature range

Transmitter body, no display

Transmitter body, with display

HMP110 probe

-40 ... +60 °C (-40 ... +140 °F)

-20 ... +60 °C (-4 ... +140 °F)

-40 ... +80 °C (-40 ... +176 °F)

Storage temperature range

-50 ... +70 °C (-58 ... +158 °F)

Electromagnetic compatibility

EN 61326-1:2006 Electrical

equipment for measurement,

control and laboratory use - EMC

requirements - Basic immunity test

requirements.

EN 55022:2006 + Am 1:2007

Information technology equipment -

Radio disturbance characteristics -

Limits and methods of

measurement. Class B.

Table 12

Inputs and Outputs

Property

Description/Value

Output signal

0 … 1 V

0 … 5 V

0 ... 10 V

User-defined between 0 ... 10 V

Min. output resistance

1 kohm

Supply voltage

10 ... 35 VDC

15 ... 35 VDC (when output

0 ... 10 V)

24 VAC (±20 %)

Current consumption at 24 VDC

8 mA, if relay closed 15 mA

Serial output

RS-485 (non-isolated)

Relay output

1 relay (max. 50 VDC, 200 mA)

Max. additional error caused by the

analog outputs after calibration at +20

°C ambient temperature

±0.1% of full scale output signal

Temperature dependence of the

analog outputs

±0.005% of full scale output signal