Selecting and scaling the analog output, Figure 17, Analog output menus – Vaisala MM70 User Manual

Page 36

Advertising
background image

USER'S GUIDE____________________________________________________________________

Selecting and Scaling the Analog
Output

0605-093

Figure 17

Analog Output Menus

To get analog measurement data, you need the analog output signal
cable, see section Options and Accessories Available on page 66. One
voltage signal channel 0...1.0V can be scaled for the selected quantity.

1.

Connect the analog output signal cable connector to the
indicator base connector. Connect the screw terminal block as
follows:
- Brown wire: Common wire (-)
- Yellow-green wire: Signal (+)

2. Open

the

MENU. First press the right arrow key, and then press

the OPEN key.

3. Select

Functions. Press the right arrow key.

4. Select

Analog output, press the right arrow key.

5. Select

Quantity and press the SET key to choose quantity.

6.

Select quantity, press the SELECT key. (Only one quantity can
be chosen at a time. Only active quantities can be selected. To
change the active quantities, enter MENU, Display, Quantities
and units
).

34 __________________________________________________________________ M210498EN-C

Advertising