Screen Setup Menu. Repeat above process for the other User Screens or press the button to return to

the Main Setup Menu. The User Screens may also be hidden if not needed.
System

The display Contrast and Auto Shutdown can be reconfigured as required. The relative humidity and

pressure sensors can also be recalibrated. Press the or buttons to highlight the appropriate selection,
and the or button to adjust or select.
The Contrast can be adjusted for better visibility depending on the ambient lighting conditions. Press the

or button to increase or decrease the contrast from 0 to 20 (0 is lightest, 20 is darkest).
The display can be set to automatically turn off in order to conserve the battery life. Auto Shutdown will
only occur after the preset time has elapsed without any button presses. Press the or button to scroll

through the Auto Shutdown options (15 minutes, 60 minutes, Off). Battery life will be shortened if the Auto

Shutdown is turned to Off.
Baro Cal - Recalibration of this sensor is not typically required, and it is not recommended that you recalibrate

without speaking to an NK technician. The pressure sensor can be calibrated if necessary. It is extremely

important to know the precise altitude and mean sea level barometric pressure at the time of calibrating

the sensor. First, set the reference altitude on the BARO measurement screen to the known altitude (see

Pressure Adjustment section for instructions). Then adjust the calibrating setting on the Baro Cal screen to

the known mean sea level barometric pressure. If recalibration is desired, the unit may also be returned to

NK for service.
Humidity Cal - Recalibration of this sensor is not typically required, and it is not recommended that you

recalibrate without speaking to an NK technician. The humidity sensor can be calibrated by “teaching” it the

correct humidity. Some special equipment is required for this calibration, including two hermetically sealed

containers and saturated salt solutions. NK offers a calibration kit, and instructions are available on

www.nkhome.com. If recalibration is desired, the unit may also be returned to NK for service.
Date & Time

The date and time, as well as date and time formats, can be adjusted. The Time Formats available are 12

hour and 24 hour. The Date formats available are day/month/year and month/day/year. (See the Date and

Time Set Up section for instructions.) Press the button to return to the Main Setup Menu.
Language

Displayed text can be set in one of five languages: English, French, German, Italian or Spanish. To choose

a language, use the or buttons to highlight the desired language. Press the button to select the

language and return to the Main Setup Menu. Otherwise, press the button to return to the Main Setup

Menu without changing languages.
Restore

Default settings for units of measure, date and time formats, and system settings can be restored. (See the

Factory Default Settings section for more information.) Press the or button to highlight the desired