Ashcroft 2089 Precision Digital Test Gauge User Manual

Page 9

Advertising
background image

– 9 –

UNITS: 12 units of measurement are available,
both English and metric, by cycling through the
UNITS key; psi, ˝Hg, ˝H

2

O (with three temperature

options, 60°F, 4°C and 20°C), ftSw, Bar, mBar,
kPa, mPa, mmHg, cmH

2

O, mmH

2

O, kg/cm

2

.

Step 1: Press the CONFIG key until the word

UNITS appears.

Step 2: Press the ENTER key.

Step 3: Press the CONFIG key once to select

ENGLISH or again to select METRIC.

Step 4: Press the ENTER key with selection of

ENGLISH or METRIC.

Step 5: Press CONFIG key to select unit of

measurement.

Step 6: Press ENTER key to finalize unit of

measurement.

UPDATE: this option allows for changing the rate
at which pressure is updated on the display
screen. The default rate measures pressure at the
maximum rate of 10* updates per second or 100
milli-seconds. Optional rates of measurement are

measured in updates per second. The options are

10*, 5, 2 or 1 update of pressure measurement
per second.

Since customer processes vary, update rates
should be selected based on the application.

To use the UPDATE option:

Step 1: Press the CONFIG key until the word

UPDATE appears.

Step 2: Press the ENTER key.

Step 3: Press the CONFIG key to select the

desired update rate.

Step 4: Press ENTER key to finalize UPDATE rate.

AUTO OFF: this option sets the amount of time
before the gauge will turn itself off after no activity.

Offerings are Never*, 2, 5, 15, or 30 minutes.

*Indicates default.