A.2 hayes at commands – Campbell Scientific COM220 56k Phone Modem User Manual

Page 22

Advertising
background image

Appendix A. Changing COM220 Settings

TABLE A-1. Selecting ME

Baud Rates via DIP Switches 3 and 4

ME Baud Rate Switch 3 Switch 4
9600

Open

Open

38400

Open

Closed

57600

Closed

Open

115200

Closed

Closed

Please note that older dataloggers (such as the CR10X, CR23X, and CR510)

cannot support the highest COM220 ME baud rates. TABLE A-2 lists

maximum datalogger baud rates.

TABLE A-2. Maximum Datalogger ME Baud Rates

Datalogger

Max Baud Rate

CR10X, CR510, CR7, CR21X

9600

CR23X

38400

CR5000

57600

CR800, CR1000, CR3000

115200

DIP Switches 5, 6, and 7: Not used

DIP Switch 8: Configuration Mode

Closing DIP switch 8 (and cycling power to the COM220) places the COM220

in configuration mode. The COM220 can be configured using an SC532 or

SC532A with a 12-Vac adapter (pn 15966) and Device Configuration Utility

version 1.6 or newer and Hayes AT commands (see below).

A.2 Hayes AT Commands

To configure the COM220 with Hayes AT commands, you will need the

Campbell Scientific Device Configuration Utility version 1.6 or newer and an

SC532 or SC532A device. An SC12 cable must be connected to the peripheral

port on the SC532 or SC532A and the CS I/O port on the COM220. Another

cable should connect the RS232 port (if using an SC532 device) or the PC port

(if using an SC532A device) to one of your computer’s serial ports. If using an

SC532A device, power will be provided to the modem by the SC532A with a

12-Vac adapter (FIGURE A-1). However, if using the older SC532, you must

apply 12 V to the COM220 via the green connector.

A-2

Advertising