5. Maintenance

The SR50A’s electrostatic transducer requires equal pressure on both sides.
Vent holes in the transducer housing are used to equalize pressure. Desiccant
is placed inside the transducer housing to prevent the possibility of condensing
humidity. The desiccant must be inspected and, if required, replaced on a
regular basis. The desiccant supplied with the SR50A uses an indicating silica
gel and will remain blue if it is still capable of absorbing moisture. Once the
desiccant becomes saturated the color will change from blue to pink. If the
SR50A is used in humid environments, the desiccant should be replaced more
frequently. To inspect or replace the desiccant, follow the procedures outlined
in Section 6 under disassembly.

The five small desiccant packets used in the transducer housing are part
number 4091.