3 cr10x example (sdi-12 “m1!”) – Campbell Scientific SR50A Distance Sensor User Manual

Page 34

Advertising
background image

SR50A Sonic Ranging Sensor


'Measure the 107 temperature sensor:

Therm107

(Air_Temp,1,1,Vx1,0,250,1.0,0)


'Use Air_Temp to calculate corrected distance:

Temp_Corr_Distance=Raw_Distance*(SQR((Air_Temp+273.15)/273.15))

'Subtract the corrected distance from the initial distance of the SR50A to the ground:

Snow_Depth=Initial_Distance

-

Temp_Corr_Distance


'Call Data Table and Store Data:

CallTable

(Table1)

NextScan
EndProg

10.1.3 CR10X Example (SDI-12 “M1!”)

;{CR10X}
;
;In this example, the SR50A is mounted 2.5 meters above the ground.
;The CR10X sends an SDI-12 command to the SR50A, which outputs a
;non-temperature compensated distance value (Raw_Dist) and signal quality value.
;Two input locations are used to store incoming data.

*Table 1 Program
01: 60

Execution

Interval

(seconds)


;Measure the 107 temperature probe:
1: Temp (107) (P11)
1: 1

Reps

2: 1

SE

Channel

;SE channel 1 used for this example

3: 1

Excite

all

reps

w/E1

;Excite channel 1 used for this example

4: 1

Loc [ T_Kelvin ]

5: 1.0

Multiplier

6: 273.15 Offset

;This converts the value to degrees Kelvin


;Use SDI-12 command "M1!" to receive Distance and Signal Quality from the SR50A
2: SDI-12 Recorder (P105)
1: 0

SDI-12

Address

2: 1

Start

Measurement

(aM1!)

3: 1

Port

;Use Ports 5-8 for CR23X

4: 3

Loc [ Raw_Dist ]

5: 1.0

Multiplier

6: 0.0

Offset


;Apply air temperature compensation to distance:
3: Z=F x 10^n (P30)
1: 273.15 F
2: 0

n,

Exponent

of

10

3: 5

Z Loc [ Ref_Temp ]


4: Z=X/Y (P38)
1: 9

X Loc [ T_Kelvin ]

2: 5

Y Loc [ Ref_Temp ]

3: 6

Z Loc [ Mult1 ]

28