Subsystem commands, Command summary, Subsystem commands -13 – Agilent Technologies B2201A User Manual

Page 139: Command summary -13

Advertising
background image

Agilent B2200 User’s Guide, Edition 2

5-13

SCPI Command Reference

Command Summary

Subsystem Commands

Command Summary

ROUT subsystem

:ROUTe subsystem has commands for controlling the signal routing.

Open a channel”: opens relays to disconnect the channel (that is, disconnects input
port from output port).

Close a channel”: closes relays to connect the channel (that is, connects input port
to output port).

Command

Description

[:ROUT]:FUNC channel_config
[:ROUT]:FUNC?

Sets the channel configuration mode.
channel_config:
ACON: Auto Config Mode
NCON: Normal Config Mode
Query returns the present channel configuration:
ACON or NCON.

[:ROUT]:CONN:RULE card_number,rule
[:ROUT]:CONN:RULE? card_number

Sets the connection rule for the specified card.
card_number: 0 or ALL for Auto Config,
1, 2, 3, 4, or ALL for Normal Config
rule: FREE (free) or SROUte (single)
Query returns the connection rule of the specified
card: FREE or SROU.
card_number: Card to check. 0 for Auto Config,
1, 2, 3, or 4 for Normal Config.

[:ROUT]:CONN:SEQ card_number,sequence
[:ROUT]:CONN:SEQ? card_number

Specifies the connection sequence mode for the
specified card.
card_number: 0 or ALL for Auto Config,
1, 2, 3, 4, or ALL for Normal Config
sequence:
NSEQ: No-Sequence mode
BBM: Break-Before-Make mode
MBBR: Make-Before-Break mode
Query returns the connections sequence mode of
the specified card: NSEQ, BBM, or MBBR.
card_number: Card to check. 0 for Auto Config,
1, 2, 3, or 4 for Normal Config.

Advertising