Agilent Technologies B2201A User Manual

Page 96

Advertising
background image

3-36

Agilent B2200 User’s Guide, Edition 2

Front Panel Operation
Setup Menus

KEY

Displays the sub menu that provides the following functions.

EXECUTE

Displays the following message.

DIAG:KEY

Press any key.

Press any front panel key and confirm the LCD display. For example, if you
press the

Port Function

key, the following message must be displayed.

DIAG:KEY

[Port Function]

Continue this for all keys. After that, press the

Enter

key twice for the normal

operation. The test result will be PASS.

If you find any wrong response, press the

Exit

key twice. The test result will be

FAIL.

Finally, press the

Exit

key to return to the previous menu.

RESULT

Displays the test result. Press the

Exit

key to return to the previous menu.

BEEPER

Displays the sub menu that provides the following functions.

EXECUTE

Starts the beeper test, and displays the following message.

DIAG:BEEP

Is beeper making 2 sounds?

Press the

Enter

key if the beeper works correctly. The test result will be recorded

as PASS.

If you find any problem, press the

Exit

key. The test result will be recorded as

FAIL.

Finally, press the

Exit

key to return to the previous menu.

RESULT

Displays the test result. Press the

Exit

key to return to the previous menu.

Advertising