Diagnostic:test:frame[:execute, Diagnostic:test:frame[:execute]? -23 – Agilent Technologies B2201A User Manual

Page 149

Advertising
background image

Agilent B2200 User’s Guide, Edition 2

5-23

SCPI Command Reference

:DIAGnostic:TEST:FRAMe[:EXECute]?

:DIAGnostic:TEST:FRAMe[:EXECute]?

This command executes the specified B2200 test, then returns the test result. See
“Selftest Menu” on page 3-34 to perform the test.

After the controller test, the B2200 status becomes same as after *RST command
execution.

Syntax

:DIAGnostic:TEST:FRAMe[:EXECute]? CONTroller | FPANel |
LED | PEN | BEEPer

Query response

test_result <newline><^END>

1: fail
0: pass

Example

OUTPUT @Agb2200;":DIAG:TEST:FRAM? CONT"

Parameter Explanation

CONTroller

Controller test

FPANel

Front panel interface test

LED

LED matrix test

PEN

Light pen test

BEEPer

Beeper test

Advertising
This manual is related to the following products: