GENERAL

INTENDED USE

Designed to meet the demanding requirements for scientific and laboratory research.

Advanced engineered design incorporates the latest in cabinet, refrigeration, temperature

control and monitoring features. Provides energy efficient, convenient, safe and reliable

performance for optimal temperature environments necessary for a wide range of plant

science, life science, pharmacy, biological, medical, clinical and industrial applications.


TECHNICAL SPECIFICATION

Voltage supply:

115V 60HZ 1 phase

Maximum fuse size: 20A

Total amp draw:

13.6A

Max set point :

44 C degrees


INSPECTION

When the equipment is received, all items should be carefully checked against the bill of

lading to insure all crates and cartons have been received. All units should be inspected for

concealed damage by uncrating the units immediately. If any damage is found, it should be

reported to the carrier at once, and a claim should be filed with the carrier. This equipment

has been inspected and tested in the manufacturing facility and has been crated in

accordance with transportation rules and guidelines. Manufacturer is not responsible for

freight loss or damage.

LOCATION

The cabinet should be located within reach of an outlet that has 115V 60HZ 20A with a

protective earth ground. The outlet should be easily accessible when installation is

complete. The refrigeration system located at the top of the cabinet requires free air

access for proper operation. Allow a minimum four-inch clearance on the top, rear, and

sides of the cabinet. The cabinet should also be leveled when it is placed in its permanent

location. Do not stack items on top of the unit. Vibration during shipping and handling may

loosen mechanical connections. Check all connections during installation. Check all

wiring, piping and fasteners. This unit has been designed for 75 deg with 50%RH at up to

2000 Meter elevation. If equipment is not used for its designed manner specified by the

manufacturer, the protection provided by the equipment may be impaired.


WARNINGS AND CAUTIONS

Do not modify cabinet construction or associated equipment assemblies.

Do not remove labeling or information supplied with the unit.


Warning: Electric Shock Hazard

.


Do Not Remove top electrical cover.

Contact a qualified service representative.