Disassembling the manifold – BrandTech Scientific Transferpette S -8/-12 User Manual

Page 45

Advertising
background image

English

45

Continued on page 46

Disassembling the Manifold

6.

Reassemble the manifold. The mani-
fold must be screwed into the handle
counter-clockwise, until it audibly
snaps into place.
Check the instrument for tightness
(see page 43, "Servicing").

5.

Push the seal on the piston with its
flat side facing upward.

(In the Transferpette

®

S -8/-12

30-300 µl unit, first replace the
additional pressure ring!)

Use the mounting tool to mount and
tighten the cleaned or new nose cone.

4.

Push the piston unit down to the

bottom. The seal will either remain
inside the nose cone or will stay on
the piston after the nose cone is
removed. Remove the seal, inspect it
and clean or replace if necessary. If
required, lubricate the piston lightly
with the supplied silicone oil.

(In the Transferpette

®

S -8/-12

30-300 µl unit, remove the additional
pressure ring from the piston for
cleaning.)