5. TECHNICAL DATA

5.2. Gauge Tube Wiring

It is strongly recommended that every customer keep a
small stock of spare gauge tubes on hand along with a
Reference Gauge Tube Calibrator.

Always try a new tube before considering circuit failures.
The Reference Gauge Tube Calibrator may be connected
to the circuit to prove whether the problem is a gauge
tube or circuit failure. If the instrument works properly with
the Reference Gauge Tube Calibrator circuit, then gauge
tube failure is a certainty.

19

5.1. Troubleshooting

The gauge tube can be checked with an ohmmeter
between pins 1 and 5 to verify that the sensor is good.
Generally, the reading is less than 2.5 ohms if good; 900
ohms if the sensor is broken.

The actual readings for each type of tube should be as
follows:

SENSOR PINS 1 TO 5

HEATER

BLACK AND GREEN WIRE

WHITE AND

APPROXIMATE RESISTANCE

RED WIRES

100 mT

2.4 Ohms

Pin 6 To 8

2000 mT

2.1 Ohms

Pin 4 To 8

20 Torr 1.3 Ohms

Pin 3 To 8