7.8.1 SETUP Display

The

MAIN -> SETUP -> DISPLAY menu contains elements that change the appearance

and function of components displayed on the main menu. Following is a description of the
elements of this menu.

Filter: The filter selection keys [LOW], [NORMAL], and [HIGH] dampen the pressure
display to reduce the affect of pneumatic noise associated with the device under test or the
test environment.

Resolution: The resolution section of the SETUP Display menus allows the user to change
the resolution of the current pressure reading to be

[4], [5] or [6] digits.

Stable Window and Delay: The stable window is the percentage of the full scale value
of the internal sensor that the current pressure can deviate (+/-) from the setpoint and still
display a stable indication. The stable delay is the number of seconds that the instrument
must remain within the stable window before the stable indication is displayed.

Language: The "VIEW" section of the SETUP display menu shows a flag, a country and a
language on a key. This is the current language. Press this key to access a menu containing
other languages that are available. "Figure - Languages" below shows the language selec-
tion screen.

Remote Status Checkbox: The Remote status check box enables or disables the remote
status icon on the main menu. This icon will show a broken wire when there is no connec-
tion to a remote computer or a connected wire if the computer is connected.

Barometer (optional): The Barometer check box enables or disables the indication of the
barometric pressure on the main menu.