Pneumatic High-Speed Pressure Controller CPC3000

WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.1

47

11498171.01 08/2009 GB

Command

Data

Response/Function

Span

desired pressure or ?

Sets span on active transducer or for
?, clears previous value, must be
> 50 %FS and has a 1 % limit.

Span?

<sp>XXXXXXX<cr><lf>

Returns span scale factor for active
transducer.

Sparity

Even, Odd, None

Sets the serial parity.

Sparity?

<sp>CCCC<cr><lf>

Returns the serial parity.

Sstop

1 or 2

Sets the serial stop bits.

Sstop?

<sp>X<cr><lf>

Returns the serial stop bits.

Stable?

<sp>(Yes or No)<cr><lf>

Returns YES if instrument is stable, or
No.

Stabledelay

0 to 65535

Sets the stable time to the number of
seconds specified.

Stabledelay?

<sp>XXXXXXX<cr><lf>

Returns the stable time.

stabletime

0 to 65535

Sets the stable time to the number of
seconds specified.

Stabletime?

<sp>XXXXXXX<cr><lf>

Returns the stable time.

StableWin

%fs value

Sets the stable window as a %FS for
the active sensor and turndown.

StableWin?

<sp>XX<cr><lf>

Returns the stable window.

Standby

None

Instrument placed in Standby Mode.

Standby?

<sp>(Yes or No)<cr><lf>

Returns Yes if instrument is in Standby,
No if otherwise.

Step

value inside upper and lower limits
and inside the range of the active
sensor and turndown.

Sets the control step size for the instru-
ment.

Step-

Optional value

Jogs the setpoint down one step.

Step+

Optional value

Jogs the setpoint up one step.

Step?

<sp>XXXXXXX<cr><lf>

Returns the control step for the instru-
ment.

Step%

Value in % of current range

Sets the control step in % of current
range.

Steppct

Value in % of current range

Sets the control step in % of current
range.

Steppct?

<sp>n.nnnnne+nn<cr><lf>

Returns the current step in % of
current range.

Units

units code or text in table below

Sets the instrument engineering units.

Units?

<sp>CCCC<cr><lf>

Returns the instrument units in a text
string.

Unlock

None

Releases Acquire locks. See Acquire?
and Release?

UpperLimit

Value inside primary xducer Range
on turndown #1 in current units.

Sets the Upper control limit for the
active transducer.

UpperLimit?

<sp>xxxxxxx<cr><lf>

Returns the upper control limit for the
active transducer.