RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 29

Advertising
background image

- 29 -

Access level

Software of moisture analyzer MA 3Y series provides four access levels:

administrator, advanced operator, operator and none (anonymous logging).

On switching on, the display is constantly active, enabling carrying out

measuring processes. These measurements, however, cannot be saved in

moisture analyzer’s database unless a user with at least minimal access level

is logged in. The minimum access level enables carrying out measuring/drying

processes and saving them in moisture analyzer’s databases using available

function.

Below table indicates access to editing user parameters, databases

and software functions depending on active access level.

Authorization

Access level

None (guest)

No access to editing all of user parameters.

it is not possible to accept a weighing record nor to start

a process.

No access to menu <Databases>.

Operator

Access to editing parameters in submenu: <Profiles;

Readout> and settings in group of parameters

<Others>, except for <Date and Time>. The operator

can start and carry out all weighing processes, and has

access to <Export weighing database to a file> and

preview data in <Databases>. An operator can define

universal variables.

Advanced

operator

Access to editing parameters in submenu: <Profiles;

Readout>; <Working modes>; <Communication>;

<Peripherals>; <Others> except for <Date and Time>.

The advanced operator can start and carry out all

weighing processes, and can erase old data from the

<Databases>.

Administrator

Access to all user parameters, functions and editing

databases.