After test completion (i.e. after testing time) the software displays window with

an on-screen keyboard for entering the temperature value as specified on the

external thermometer of the temperature adjustment kit.

Enter and accept the temperature value. The software displays a window on test

summary.

In the test summary window, the user can order printing a report from drying

chamber test. On selecting one of the available options the software

automatically returns to displaying the menu window: <Adjustment>.