Front panel procedure: (P-band waveguide example)
1. Prior to modifying or generating a cal kit, store
one or both of the cal kits in the 8510’s non-
volatile memory to a disk.

2. Select CAL menu, MORE.

3. Prepare to modify cal kit 2: press MODIFY 2.

4. To define a standard: press DEFINE STANDARD.

5. Enable standard no. 1 to be modified: press 1,
X1.

6. Select standard type: SHORT.

7. Specify an offset: SPECIFY OFFSETS.

8. Enter the delay from Table 1: OFFSET DELAY,
0.0108309, ns.

9. Enter the loss from Table 1: OFFSET LOSS, 0,
X1.

10. Enter the Z

0

from Table 1: OFFSET Z

0

, 50, X1.

11. Enter the lower cutoff frequency: MINIMUM
FREQUENCY, 9.487 GHz.

12. Enter the upper frequency: MAXIMUM FRE-
QUENCY, 18.97 GHz.

13. Select WAVEGUIDE.

14. Prepare to label the new standard: PRIOR
MENU, LABEL STANDARD, ERASE TITLE.

15. Enter PSHORT 1 by using the knob, SELECT
LETTER soft key and SPACE soft key.

16. Complete the title entry by pressing TITLE
DONE.

17. Complete the standard modification by press-
ing STANDARD DONE (DEFINED).

Standard #1 has now been defined for a

1

/

8

λ P-band

waveguide offset short. To define the remaining
standards, refer to Table 1 and repeat steps 4 -17.
To define standard #3, a matched load, specify
“fixed.”

The front panel procedure to implement the class
assignments of Table 2 for the P-band waveguide
cal kit are as follows:

1. Prepare to specify a class: SPECIFY CLASS.

2. Select standard class S

11

A.

3. Inform the 8510 to use standard no. 1 for the
S

11

A class of calibration: l, X1, CLASS DONE

(SPECIFIED).