Agilent Technologies 8510 User Manual

Page 25

Advertising
background image

25

4. Change the class label for S

11

A: LABEL CLASS,

S

11

A, ERASE TITLE.

5. Enter the label of PSHORT 1 by using the knob,
the SELECT soft key and the SPACE soft key.

6. Complete the label entry procedure: TITLE
DONE, LABEL DONE.

Follow a similar procedure to enter the remaining
standard classes and labels as shown in the table
below:

Finally, change the cal kit label as follows:

1. Press LABEL KIT, ERASE TITLE.

2. Enter the title “P BAND.”

3. Press TITLE DONE, KIT DONE (MODIFIED). The
message “CAL KIT SAVED” should appear.

This completes the entire cal kit modification for
front panel entry. An example of programmed
modification over the GPIB bus through an exter-
nal controller is shown in the Introduction To
Programming section of the Operating and Service
manual (Section III).

Standard

Standard

Class

class numbers

label

S

11

B

2

PSHORT 2

S

11

C

3

PLOAD

S

22

A

1

PSHORT 1

S

22

B

2

PSHORT 2

S

22

C

3

PLOAD

FWD TRANS

4

THRU

FWD MATCH

4

THRU

REV TRANS

4

THRU

REV MATCH

4

THRU

RESPONSE

1,2,4

RESPONSE

Advertising