Calibration Procedure:

NOTE: The Triton is designed with a recessed CAL key so that you won’t press it by accident during
normal use. The CAL key is located between the TARE and ON/OFF keys. There is a small hole in
between those keys - use a bent paperclip to press the recessed CAL key in that hole.

1) Turn the scale OFF and Place the it on a Flat, very stable surface

(Be certain the scale is OFF)

2) Use a bent paperclip to press and hold the hidden CAL key for 5 seconds until a random large number
appears on the display - then release the key.

3) Press the CAL key again, the display will show 2Ero and then

100.0

4) Gently place a 100 gram weight on the scale and wait 3 seconds.

5) Press the CAL key again. The display will show CAL and then PASS, calibration is complete. Gently
remove the weight and turn the scale Off.

NOTE: if after calibration your scale does not read accurately, this indicates

calibration error and the calibration process should be repeated more slowly.

Please calibrate on a very stable flat surface

INACCURACY / ERROR___________________________________

The primary reasons for inaccuracy or malfunction are low batteries, incorrect calibration, overload or
operating on an unstable surface. Please keep this in mind and maintain and operate your scale
properly. The scale is a precise instrument and must be handled with the utmost care and caution.

* Power Up Segment Test

When first turning the unit on, all segments of the display will appear as shown below. This display will
remain for approximately 2 seconds and then change to 0. 8:8:8:8

KEY PAD FUNCTIONS

* ON/OFF

Press this key to turn unit on or off.

* TARE

Press tare to reset the scale to zero. Tare can be used for eliminating the weight value of a container or
the integrated sliding tray.

Advertising