Alerts – BACtrack S30 User Manual

Page 6

Advertising
background image

8

9

STEP 2: Start-up

when the countdown approaches 1,

begin to inhale a deep breath.

STEP 3: Breath Sampling

when the countdown reaches zero, the

BaCTraCK will begin beeping and the

lCD segments will circle. Blow hard,

continuously, for five full seconds. while

a user is blowing, the BaCTraCK will

sound a continuous tone. Keep blowing

until two beeps are heard.

STEP 4: Breath Analysis

when finished taking a breath sample,

the BaCTraCK will circle the outer lCD

fields for three seconds while analyzing

the breath sample.

STEP 5: Display %BAC

Once the sensor technology has

analyzed the breath sample, the

estimated BaC (Blood alcohol Content)

value is flashed on the screen.

The test result will be displayed for 10

seconds. Then the BaCTraCK will beep

five times before powering off.

Alerts

Battery Low

The battery power is low. Please

replace the batteries.

Flow Error

The breath sample was not sufficient.

Please repeat the testing process,

ensuring a continuous, five second

breath sample.

Time Out

If the user does not blow within 10

seconds, the unit displays [err] and

shuts down.

Calibration

The unit requires recalibration or

maintenance. Please refer to the

Verification and Calibration section

of this manual.