Chapter – Advantech ADAM-5000 User Manual

Page 69

Advertising
background image

Chapter

6

6-7

6.2 ADAM-5013 RTD Input Resistance Calibration

1. Apply power to the module and let it warm up for about 30 minutes.

2. Make sure that the module is correctly installed and is properly

con- figured for the input range you want to calibrate. You can
use the ADAM utility software to help in this.

3. Connect the correct reference self resistance between the screw

terminals of the ADAM-5013 as shown in the following wiring
diagram. Table 2 below shows the correct values of the span and
zero calibration resistances to be connected. Reference
resistances used can be from a precision resistance decade box or
from discrete resistors with the values 60, 140, 200 and 440 ohms.

Figure 6-5: Applying calibration resistance

4. First, with the correct zero (offset) calibration resistance connected

as shown above, issue a Zero Calibration command to the module
using the Calibrate option in the ADAM utility software.

5. Second, with the correct span resistance connected as shown

above, issue a Span Calibration command to the module using
the Calibrate option in the ADAM utility software. Note that the
module zero calibration must be completed prior to the span
calibration.

Note: If the above procedure is ineffective, the user must first issue an

RTD Self Calibration command $aaSi2 to the module and then
complete steps 4 and 5 after self calibration is complete.





Advertising