Solving pr oblems, Err or codes and symbols – Bayer Contour-TS User Manual

Page 27

Advertising
background image

25

Solving

Pr

oblems

www.bayerdiabetes.com/us

WHAT YOU SEE

WHAT IT MEANS

WHAT YOU SHOULD DO

Some display seg-
ments do not appear
when the meter
initially turns on.

Possible meter
electronics failure.

Press and hold while the
meter is off. Compare meter
display with the picture on
page 3. If the display is still
not correct, call Customer
Service.

Control test result
is out-of-range (too
high or too low).

Test strip is

deteriorated due
to exposure to
humidity or heat.

Control solution

is past expiration
date or is past
the open use
date.

Test strip is past

expiration date
or is past the
open use date.

Incorrect control

solution was
used.

Control solution

is not at room
temperature.

Control solution

is not mixed
well.

Run another control test

with a new test strip.

Check all expiration dates.

Do not use expired testing
materials.

Make sure you are using

C

ONTOUR

TS control

solution.

Wait 20 minutes until

control solution is at room
temperature. Try again.

Gently rock the control

bottle to ensure the con-
trol solution is mixed well.

If result is still out-of-

range, retest with a new
test strip and control
solution.

If result is still out-of-

range, call Customer
Service.

Diabetes Care Customer Service at 1-800-348-8100.