2.4.3 DER Indication Mode

In the DER mode (Figure 2.1) the following values appear on the LCD:

DER (µR/h, mR/h, R/h);

DER on the analog scale in a logarithmic gauge (seven segments);

Coeffi cient of variation in percents;

Averaging time of DER values (Range of the averaging time

indication is from 1 up to 2999 s. If the averaging time exceeds

2999 s, the symbols - - - appear on the LCD).

If the measured DER value is above the upper limit of the DER indication

range, the LCD will show the overload symbol “OL” and a noncontinuous

audible signal will sound.

In the DER mode the collected statistics of DER measurement can be

cleared and measurement reactivated by simultaneously pressing the

LIGHT + MODE buttons.

2.4.4 DE Indication Mode

In the DE mode (Figure 2.1) the following values are indicated on the LCD:

DE (µR, mR, R);

DE accumulation time.

DE and DE accumulation time can be reset in the set mode by

simultaneously pressing the LIGHT+MODE buttons (Figure 2.2).

In data transmission to PC mode it is possible to set an inhibit for DE

reset using the buttons. There are two methods of measurement of the DE

accumulated over a certain period of time.

The fi rst method (recommended). In the beginning of measurement the

DE should be reset using the keys or the PC and the program “System

of date collection and processing for the 1621 device.” Then the dose

displayed at the end of the period will be the dose accumulated over the

period of measurement.